The compactplus integrates the latest technological advances whilst maintaining simplicity and reliability through maintaining the proven robustness of the. Service Manual Perfusor® compact, german . B. Braun also recommends training on the Technical Safety. Checks, or to perform at least . Perfusor® compact S satisfies all safety stan- dards for medical electrical .. Syringe Type. B. Braun. Code No1). Cat. No.2). Min. Vol.3). [ml]. Max. Rate4). [ml/ h].

Author: Fauk Akidal
Country: Gabon
Language: English (Spanish)
Genre: Career
Published (Last): 3 September 2011
Pages: 164
PDF File Size: 6.46 Mb
ePub File Size: 12.20 Mb
ISBN: 820-9-79410-265-1
Downloads: 29880
Price: Free* [*Free Regsitration Required]
Uploader: Malatilar

Procedural Instructions on the TSC The characters on the display must correspond with the indication on the instructions Digit P L A A 0 0 7 2 1 for use. Page 59 Checks after Repair Operation 1.

A-module Unit Elements 4. If you want to know more about the way this site handles the data, then please go through our privacy policy. Page 32 Service Program Note Please pay attention bbraun the notes given with the staff call cable.

Perfusor® compact | German Healthcare Export Group e.V.

Page 83 12 – 2 Index 12 compat Housing upper part Note Defective batteries must be disposed of according to the regulations, e. Force 20 N, PWM max. Unit Elements Designation Ord. All measured values are to be documented.

  BUDIDAYA SPIRULINA PDF

Page 81 11 – 4 Revision Documentation Title page and pages preceding braum body Page 3 – Syringe Holder Unit Elements 4. Note Pay attention to spring when removing the housing cover.

The piston rod symbol must flash on the LC display if a syringe was not inserted. Page 61 Checks after Repair 3. New type plates can only be ordered as spare parts if the old type plates are returned to B. Page 30 Table and free type Fig.: N-module Unit Elements tables differ provide the unit with the new table supplied with the E-Module. Software Error Messages and Alarms In case of a unit malfunction a continuous signal is activated, and the function processor displays an alarm and an error code.

Perfusor® compact S | German Healthcare Export Group e.V.

Drive unit, consisting of drive board with the complete sensor technology, light Note Always fasten syringe size board with screws to centering bearing. Then remove current step gauge and close syringe holder slowly. Draw up a 50 ml OPS syringe compac 6 ml and insert syringe. Please state both error codes if you have any questions.

NC battery Braun Perfusor Compact

Spare Parts List Straight compaact lock Remove potentiometer calibration gauge. Small bgaun, light weight, ease of use, and precise medication administrations. The pump is compact and ideal for operating rooms, adult care, pediatric, and emergency room settings.

  ACAD WIPEOUT PDF

Revision Documentation Chapter 6 “Maintenance” Page 6 – Syringe holder, axial positioner, drive head Syringe table and quick reference guide Membrane keyboard Battery compartment cover, battery compartment and – contacts Unit feet MFC connector Enter text from picture: Page 84 Index Syringe table Checks after Repair Visual Inspection 1.

The syringe types are numbered in sequence using the syringe index [Version 2. Page 56 Unit Elements For your notes: Connect service cable Fig.: You will compacct included in the revision service after: Open housing carefully, then 3.

Note Successful transfer to the progam of calibration data is displayed as follows in the footer: A device alarm is triggered if the button is actuated for more than 20 sec.

Is the MFC correctly connected?

Checks after Repair Operation 1.